Leynow specialises in laboratory oil-free vacuum pumps: cost-effective, full range, Maintenance-free and easy to operate. Small Oil-free Vacuum Pumps feature an oil-free design, require no maintenance, are compact and light, achieve high Vacuum Levels, run quietly and come with a quality guarantee.
An oil-free Vacuum Pump can reach an ultimate vacuum of 2000-4000 Pa, and up to 270-670 Pa when combined with an atmospheric ejector. A water ring pump can also be used as a Compressor, in which case it is called a water ring compressor; it is a low-pressure compressor with a pressure range of 1-2x105 Pa gauge. To meet the process targets, efficiency and service life requirements of different products, different Vacuum ranges require different Vacuum System configurations.
This product mainly has two rotors rotating in opposite directions inside the pump body, driven by a pair of timing gears. The rotors mesh with each other inside the pump body but retain a clearance, the size of which depends mainly on the liquid viscosity; above a certain viscosity range the clearance must be increased accordingly.
When the motor rotor turns, part of the volume at the inlet of the pumping chamber expands and draws gas in through the inlet port; the gas passes through the air filter and enters the pump chamber through the open anti-return oil valve. As the motor rotor turns further, the vanes separate part of the pump chamber from the inlet. These chamber volumes decrease and the gas is compressed. When the gas is slightly above Atmospheric Pressure it is discharged from the pump chamber through the automatic exhaust valve.
So what causes an oil-free vacuum pump to fail to start normally, and how can it be solved?
Possible causes:
1. The motor starter overload protection is set too low or the setting value is too low.
2. If an AC motor is fitted: the motor capacitor is damaged.
3. Check the condition of the pump oil - oil return, oil temperature too low or deteriorated oil can fill the pump chamber with oil. Observe whether the oil has emulsified (white foam) or carbonised (turned black); deteriorated oil affects the friction between the vanes and the rotor and can cause the vanes to seize in the pump chamber.
Solutions:
1. Repair or replace the motor.
2. Compare the overload protection setting of the motor starter with the data on the motor nameplate and correct it if necessary. If the ambient temperature is high, the setting value can be raised by 5% above the normal motor current.
3. Replace the pump oil, disassemble and repair the cylinder and other components, and replace the pump vanes if they are seriously damaged.